TMX MAINFRAME | |
MAINFRAME CHASSIS | |
Maximum Analog Modules |
3 (6 with optional expansion unit) |
Maximum Analog Waveforms | 48 (96 with optional expansion unit) |
Event Inputs (TTL) | 16 |
Derived Channels | Addition, Subtraction, Multiplication, Division, Exponential, Sine, Cosine, Tangent, Square Root & Absolute Value |
Pre-capture filters | Low pass, high pass, band pass, band stop, RMS |
Post-capture filters | Low pass, high pass, band pass, band stop |
User Engineering Units | Yes |
Calibration | Semi-automated to external reference |
DATA ACQUISITION RECORDING | |
Operational Modes |
Scope, Review, Real-time (strip-chart) |
Recording Method | Internal removable 1 Tbyte SATA disk drive |
Total Capacity | Over 400 billion samples |
Maximum Record | Up to the drive capacity |
Data Stored | Raw (unfiltered) data saved to the drive |
Time Stamp | Time and date automatically saved with data |
Trigger Types | Edge, window, slew, event, manual & external |
Trigger Point | Amount of pre and post trigger is user adjustable |
Auto Re-Arm | Automatic stacking of captures |
File Information | Information on units, measurement range and sample rate saved with data |
COLOR DISPLAY | |
Type |
Active matrix color LCD (TFT) |
Viewing Area | 17 inch / 43.18 cm (diagonal) |
Resolution | 1280 x 1024 |
Touch | Full screen, resistive |
COMPLIANCE / ENVIRONMENTAL | |
Safety |
EN 61010-1, 2nd Edition (2001) , UL 61010-1:2004, 2nd Edition CAN/CSA C22.2 No. 61010-1:2004 2nd Edition |
EMC | FCC Part 15, Subpart B, Class A; EN 61326-1 Class A |
Power Harmonics | IEC 1000-3-2 |
Operating Temp | 40 to 105 °F (5 to 40 °C) |
Operating Humidity | 10 % to 90 % non condensing |
Shock | MIL-STD-810F Method 516.5, Procedure I |
Vibration | MIL-STD-810F Method 514.5, Procedure I |
POWER | |
AC Input Voltage Range |
102 to 264 VAC |
AC Input Voltage Frequency Range | 47 Hz to 63 Hz |
Power Consumption | 300 W maximum |
DC Input Voltage | 24 VDC (Regulated +/- 0.5V) @ 11A |
DC Input Power | 264 W maximum |
PHYSICAL | |
Enclosure |
Aluminum, with armored end caps |
Dimensions | 14.5” ( 36.8 cm) H x 19” ( 48.3 cm) W x 7.5” (19.1 cm) D (without handle) |
Weight (including 3 modules) | 37 lbs (15.28 kg) varies based on installed modules |
INTERFACE | |
Ethernet |
1000BaseT |
VGA | For displaying data on an external monitor |
USB 2.0 (8 ports per unit) | For external peripherals and file export |
Expansion Port | For connection of optional TMX-E expansion chassis |
TMX - INPUT MODULE SPECIFICATIONS | |
UNIV-6 Universal Isolated Voltage Module with DC Bridge | |
UNIV-6 General Specifications | |
Channels (per module) |
6 |
Maximum Sample Rate (per channel) | 800 kHz (400 kHz with optional TMX-E expansion chassis) |
Maximum Bandwidth (Single Ended Input) | 100 kHz (-3db) (400 V, 200 V and 50 V attenuators) 90 kHz (-3db) (10 V and 1 V attenuators) |
Maximum Bandwidth (Differential Input) | 50 kHz (all attenuators) |
Isolation | 250 Vrms or DC, Cat II (channel to chassis, channel to channel) |
A/D Resolution | 16 bit (separate A/D per channel) |
Anti-Aliasing Filter | 4 pole Bessel |
Accuracy (25°C) | ± 0.07 % of attenuator |
Frequency Counter | 1 (first channel of each UNIV-6 module) |
Frequency Counter Modes | Frequency to Voltage (time and cycle based), Duty Cycle , Pulse Width and Period Detector |
Frequency Counter Range | 2 Hz - 40 kHz (48 Hz - 100 kHz for cycle based frequency counter) |
Frequency Counter Accuracy | ± 0.07 % of measurement + .002 Hz |
Minimum Counter Input Amplitude | 25 % of span for frequency and pulse counters 90 % of span for all other counter modes |
Pulse Counter Range | 64000000 maximum (16 bit display resolution) |
Pulse Width Range | 10 uS - 40000 uS |
Pulse Width Accuracy | 0.7 uS + .00167 % of span |
Period Width Range | 5 uS - 90000 uS (11 Hz - 200 kHz) |
Period Width Accuracy | 0.02 % of measurement + 0.00167 % of span with a maximum of 1.0 uS |
Duty Cycle Range | 15 hz - 10 kHz range with 20 % - 80 % duty cycles |
Duty Cycle Accuracy | 0.5 % of measurement |
Cold Start Drift | < 0.1 % of attenuator (60 min.) |
UNIV-6 Single Ended Voltage Input | |
Input Type |
Single Ended, AC/DC coupled |
Isolated Input Connector | Guarded banana jacks (red/black) |
Max Rated Input | 250 Vrms or DC, Cat II |
Max Transient Input | ± 800 V (not to exceed 250 Vrms) |
Specified Ranges | 200 mVFS to 800 VFS |
Attenuators | 1 V, 10 V, 50 V, 200 V, 400 V |
Minimum Input Impedance | 1 MΩ |
Intrinsic Noise (pk-pk) | < 0.17 % of attenuator + 0.07 % of span (1 V attenuator) < 0.08 % of attenuator + 0.08 % of span (10 V through 40 V attenuators) |
IMR @ 60 Hz | > -75 db |
UNIV-6 Differential Voltage Input / Bridge Measurements | |
Input Type |
Differential, DC coupled |
Isolated Input Connector | Screw terminal header (removable) |
Max Rated Input | ± 3 VDC |
Max Transient Input | ± 20 VDC |
Specified Ranges | 5 mVFS to 2 VFS |
Attenuators | 50 mV , 500 mV, 1000 mV |
Minimum Input Impedance | 300 kΩ (150 kΩ balanced to isolated common) |
Intrinsic Noise (pk-pk) | < 0.14% of attenuator + 0.08% of span (50 mV attenuator) < 0.08 % of attenuator + 0.09 % of span (500 mV & 1000 mV attenuators) |
CMR @ 60 Hz | > -85 db |
Excitation | DC Adjustable 0.1 to 10V @ 30 mA maximum |
NIDV-16 Non-isolated Differential Voltage Module | |
Channels (per module) |
16 |
Maximum Sample Rate (per channel) | 200 kHz (100 kHz with optional expansion chassis) |
Maximum Bandwidth | 40 kHz |
Input Type | Differential, non-isolated DC coupled |
Differential Input Connector | 25-pin D-sub male connectors (8 channels per connector) |
A/D Resolution | 16 bit (separate A/D per channel) |
Anti-Aliasing Filter | 4 pole Bessel |
Maximum Rated Input | ± 50 VDC (35 Vrms) |
Maximum Transient Input | ± 60 V Peak |
Specified Ranges | 80 mVFS to 100 VFS (with zero offset) |
Attenuators | 200 mV, 400 mV, 800 mV, 2 V, 4 V, 5 V, 10 V, 25 V, 50 V |
Minimum Input Impedance | > 500 kΩ (250 kΩ balanced to signal common) |
Accuracy (25°C) | ± 0.07 % of attenuator (800 mV, 2, 4, 5, 10, 25 & 50 V attenuators) ± 0.1 % of attenuator (400 mV attenuator) ± 0.15 % of attenuator (200 mV attenuator) |
Intrinsic Noise (pk-pk) | < 0.1% of attenuator + 4 mV |
CMR @ 60 Hz | > -60 db |
Auxiliary Power Output | 12V @ 200 mA (using pin 25 on both connectors) |
Frequency Counter | 2 (first and second channel of each NIDV-16 module) |
Frequency Counter Modes | Frequency to Voltage (time and cycle based), Pulse Counter, Quadrature Event Counter, Duty Cycle , Pulse Width and Period Detector |
Frequency Counter Range | 2 Hz - 40 kHz (48 Hz - 100 kHz for cycle based frequency counter) |
Frequency Counter Accuracy | ± 0.05 % of measurement + .002 Hz |
Minimum Counter Input Amplitude | 25 % of span for frequency and pulse counters 90 % of span for all other counter modes |
Pulse Counter Range | 64000000 maximum (16 bit display resolution) |
Pulse Width Range | 40 uS - 40000 uS |
Pulse Width Accuracy | 1.5 uS + .00167 % of span |
Period Width Range | 5 uS - 90000 uS (11 Hz - 200 kHz) |
Period Width Accuracy | 0.2 % of measurement + .00167 % of span |
Duty Cycle Range Accuracy | 0.5% (Inputs in the 15 Hz - 10 kHz range with 20% - 80% duty cycles) |
Duty Cycle Accuracy | 0.5 % of measurement |
IHVM-6 Isolated High Voltage Module | |
Channels (per module) |
6 |
Maximum Sample Rate (per channel) | 800 kHz (400 kHz with optional TMX-E expansion chassis) |
Maximum Bandwidth |
60 kHz (-3db) |
Input Type | Differential, DC coupled |
Isolated Input Connector | Guarded banana jacks (red/black) |
A/D Resolution | 16 bit (separate A/D per channel) |
Anti-Aliasing Filter | 4 pole Bessel |
Max Rated Input |
600 Vrms or 1000 VDC, Cat IV |
Max Transient Input | +/- 2000 V peak |
Specified Ranges | 1 VFS to 2000 VFS |
Attenuators | 40 V, 200 V and 1000 V |
Minimum Input Impedance | 4 MΩ (2 MΩ balanced to isolated common) |
Accuracy (25°C) | ± 0.07 % of attenuator |
Intrinsic Noise (pk-pk) | 40 V attenuator (< 0.11 % of attenuator + 0.08 % of span) 200 V attenuator (< 0.04 % of attenuator + 0.08 % of span) 1000 V attenuator (< 0.02 % of attenuator + 0.08 % of span) |
IMR @ 60 Hz | -65 db (40 V, 200 V attenuators) -60 db (1000 V attenuator) |
Frequency Counter | 1 (first channel of each IHVM-6 module) |
Frequency Counter Modes | Frequency to Voltage (time and cycle based), Pulse Counter, Duty Cycle, Pulse Width and Period Detector |
Frequency Counter Range | 2Hz - 60 kHz |
Frequency Counter Accuracy | ± 0.07% of measurement + .002 Hz |
ITCU-12 Isolated Thermocouple Module | |
Channels (per module) | 12 |
Input Type | Type U miniature thermocouple (12 connectors) |
Isolation | 250 Vrms or DC, Cat II (channel to chassis, channel to channel) |
Maximum Bandwidth | 5 Hz update rate (TC sampled at 2.5 Hz) |
Max Rated Input | ± 10 V |
A/D | 24 bit Sigma Delta (separate A/D per channel) |
Anti-Aliasing Filter | Inherent |
Resolution | 0.1 °C |
Thermocouple Types | J, K, E, T, N, B, R, S & C |
Specified Range (J) -210 to 1200 °C | Accuracy (25°C) J (-200 to 0) +/- 3.0 °C (0 to 1200) +/- 1.0 °C |
Specified Range (E) -200 to 1000 °C | Accuracy (25°C) E (-200 to -100) +/- 2.0 °C |
Specified Range (N) -200 to 1300 °C | Accuracy (25°C) N (-200 to -50) +/- 1.5 °C (-50 to 1300) +/- 0.8 °C |
Specified Range (R) -20 to 1768 °C | Accuracy (25°C) R +/- 4.0 °C |
Specified Range (K) -200 to 1372 °C | Accuracy (25°C) K (-200 to 0) +/- 2.5 °C (0 to 1372) +/- 1.5 °C |
Specified Range (T) -200 to 400 °C | Accuracy (25°C) T (-200 to -100) +/- 3.0 °C (-100 to 400) +/- 1.0 °C |
Specified Range (B) 600 to 1820 °C | Accuracy (25°C) B +/- 4.5 °C |
Specified Range (S) -20 to 1768 °C | Accuracy (25°C) S +/- 4.0 °C |
Specified Range (C) 0 to 2316 °C | Accuracy (25°C) C +/- 2.5 °C |
Specified Range 100 mV | Accuracy (25°C) +/- 0.05 % of attenuator |
Cold Junction Compensation | Yes, on each channel |
Compensation Error | Included in above accuracy specifications |
Intrinsic Noise (pk-pk, J,N) | < 0.15 °C |
Intrinsic Noise (pk-pk, K,E,T) | < 0.10 °C |
Intrinsic Noise (pk-pk, B,R,S) | < 1.0 °C |
Intrinsic Noise (pk-pk, C) | < 0.5 °C |
Intrinsic Noise (pk-pk, 100 mV) | < 0.007 % of attenuator |
IMR | > 110 dB @ DC |
Linearization | NIST ITS-90 |
Note: specified accuracy does not include probe errors. | |
COMING SOON (please contact Test & Measurement Sales for additional details) | |
IBRM-6 Isolated Bridge Module | |
Channels (per module) | 6 |
Maximum Sample Rate (per channel) | 800 kHz (400 kHz with optional TMX‐E expansion chassis) |
Maximum Bandwidth | 50 kHz |
Input Type | Isolated Differential |
Isolation | 250 Vrms or DC, Cat II |
TEDS Capability | YES |
IEPE-6 Isolated Piezoelectric Sensor Module | |
Channels (per module) | 6 |
Maximum Sample Rate (per channel) | 800 kHz (400 kHz with optional TMX‐E expansion chassis) |
Maximum Bandwidth | Up to 30 kHz |
Input Type | Isolated Differential |
Isolation | 250 Vrms or DC, Cat II |
TEDS Capability | YES |
TMX OPTIONS | |
TMX-VA Video / Audio Acquisition | |
Analog Input Type / Connector | Composite / BNC |
Supported Video Formats | NTSC, PAL |
Synchronization with waveforms | +/- 1 video frame |
Maximum Capture Length | NTSC: Limited to remaining data capture drive space PAL: Limited to 2 hour data captures |
Supported TMX Operating Modes | Realtime |
LookBack Capability | No |
Pre-trigger capture | No |
Trigger capability | Manual Trigger Only |
NTSC Capture Resolution | 720 x 480 pixels |
NTSC Capture Rates | 5, 10, 20, 30 fps (frames per second) |
PAL Capture Resolution | 720 x 576 pixels |
PAL Capture Rates | 5, 25 fps (frames per second) |
Multimedia Format | AVI with MPEG-4 compression |
Audio Resolution | 16-bit |
Audio Mode | Mono |
Audio Capture Rate | 11 kHz |
TMX-E Expansion Chassis (Requires TMX mainframe chassis for operation) | |
Maximum Analog Modules |
3 |
Maximum Analog Waveforms | 48 |
Enclosure | Aluminum, with armored end caps |
Dimensions | 14.5” (36.8 cm) H x 19” (48.3 cm) W x 5.04” (12.8 cm) D (without handle) |
Weight (including 3 modules) | 15 lbs (6.8 kg) varies based on installed modules |
Input Voltage Range | 102 to 264 VAC |
Frequency Range | 47 Hz to 63 Hz |
TMX-R Rackmount Version* (Fits standard 19" racks) | |
Maximum Analog Modules** |
6 |
Maximum Analog Waveforms | 96 |
Enclosure | Aluminum, rackmountable |
Dimensions | 15.75” (40 cm) H x 19” (48.2 cm) W x 17.15” (43.6 cm) D (without handle) |
Weight (including 6 modules) | 47 lbs (21.3 kg) varies based on installed modules |
Input Voltage Range | 102 to 264 VAC |
Frequency Range | 47 Hz to 63 Hz |
Ethernet | 1000BaseT |
VGA | For displaying data on an external monitor |
USB 2.0 (9 ports: 8 in back and 1 in front of unit ) | For external peripherals and file export |
* The DATA ACQUISITION RECORDING and COLOR DISPLAY sections above also apply to the TMX-R rackmount version. | |
**When more than three input boards are used for data acquisition or monitoring in the TMX-R rackmount system, the maximum sample rate per channel will be one half the specified maximum sample rate for a specific input board. Maximum bandwidth per channel is not impacted. | |
Specifications are subject to change. Registered trademarks belong to their respective companies. |