The Ntron Model 5100 provides an affordable and versatile Oxygen Analyzer solution ideal for portable and fixed gas generation standard process monitoring. The unit delivers high-purity gas measurement in a range from 0-1% through to 0-100% in an auto range or fixed range modes.
A small footprint and compact size makes the Ntron 5100 easy to install. The unit’s expected operational life is five to ten years.
The low-maintenance, long-life Zirconia sensor has virtually unlimited shelf life and will not dry out or freeze.
Specifications:
Sensor: Mini Zirconia Sensor
Measurement Range: 0-1%/0-10%/0-25%/0-50%/0-100% auto range or fixed range
Display: 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN
Response Time: <15 sec. for 90% step change
Intrinsic Error (accuracy):
At constant temperature: ±2% range
Over operating temperature: ±5% range
When calibrated @ target concentration @ STP: ±1% range
Operating Temperature: 5 to 45º C
Signal Interface: RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC or 0-5VDC or 0-10VDC analog oxygen signal
Alarm Outputs: Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay.
Sample Port: 1/8” NPT female ports
Electrical Connections: Removable terminal blocks on back of analyzer
Power, Analyzer: 5W at 115/230VAC 50/60Hz, single phase or 10 Watts @ 12-30VDC (battery backup capability)
Environmental Ratings, Analyzer:
Faceplate: NEMA type 4, IP66
Housing: NEMA type 1, IP20
Warranty: Electronics and sensor: One yr.
Weight: Analyzer: Less than 2 lb. (0.9 kg)