6540阻抗分析仪6530/6520介绍
PRODUCT FEATURES |
MEASUREMENT PARAMETERS | |
.Capacitance (C) | .Dissipation Factor (D) |
.Inductance (L) | .Quality Factor (Q) |
.Resistance (R) | .Impedance (Z) |
.Reactance (X) | .Admittance (Y) |
.Conductance (G) | .Phase Angle (Ø) |
.Susceptance (B) |
Technical Specifications 6540A
Measurement parameters
Any of the following parameters can be measured displayed :
AC Functions
Capacitance (C)
Inductance (L)
Resistance (R)
Reactance (X)
Conductance (G)
Susceptance (B)
Dissipation Factor (D)
Quality Factor (Q)
Impedance (Z)
Admittance (Y)
Phase Angle (Ø)
Display format
Series or parallel equivalent circuit.
Test conditions
Frequency range
1kHz to 120 MHz
Accuracy of set frequency ±0.005%
AC Drive level
10 mV to 1V rms*
200µA to 20 mA rms*
* Varies with frequency
Signal source impedance : 50nominal
DC bias current (optional)
10 mA of DC bias current
40V of DC bias voltage
Mode of operation
Graphical sweep mode
Allows graphical sweep of any two measurement parameters
Swept parameters : -frequency, drive level or DC bias
Meter mode
Allows the instrument to be used as a stard LCR meter.
Setup data
up to 20 instrument setups may be stored locally for each mode.
Measurement connections
Four front panel BNC connectors permit three or four terminal connections with the screens at ground potential.
Measurement accuracy
Dissipation factor
±0.0005 (1+D2)*
Quality factor
±0.05% (Q+1/Q)*
*Varies with frequency, drive level measured impedance.
General
Power supply
Input voltage 90V AC to 264 V AC (Actoranging)
Mains frequency
50~60 Hz
Display
8.4 VGA (640 X 480) color TFT with touch panel
Local printer
HP-PCL compatible graphics printing
Epson compatible text / ticket printing
Network printer
HP-PCL compatible graphics printing